The cold room display reads 5°C. Are the cartons near the door, along the walls, and on the top shelf all staying within 2–8°C?
Temperature mapping uses data loggers at multiple locations to find the warmest and coldest points and identify areas with the largest fluctuations. The results help quality teams determine which storage areas meet the required temperature range under the conditions tested.
How to Map a 2–8°C Cold Room
1. Define the study. Set the acceptance criteria based on the products’ storage requirements. Specify the test conditions, sampling interval, and study duration. Record events such as door openings, restocking, and defrost cycles.
2. Place calibrated loggers at selected points. Use a risk assessment to choose locations, including different shelf heights, corners, and areas near doors or ventilation. Mark each logger ID, location, and height on the layout.
3. Compare the records in FRESVUE LoggerAnalysis. Check that the timestamps align and the records are complete, then compare reports from multiple loggers in the same workspace. If the door area warms up after opening, compare its curve with those from the middle and upper shelves to see how far the warming reached and how quickly each point recovered.
Figure 1. LoggerAnalysis interface showing the temperature curve, device information, and summary statistics for one logger.
Which FRESVUE Logger Fits Your Study?
ZTAG-T Single-Use USB Temperature Data Logger — For a one-off study or when retrieving devices for reuse is impractical. Product details ZLOG-T Reusable USB Temperature Data Logger — For periodic requalification or projects that rotate the same devices across several cold rooms. Product details ZLOGPRO-30XT High-Accuracy USB Temperature Data Logger — A reusable logger with specified accuracy of ±0.3°C, suited to studies with tighter measurement accuracy requirements. Product details For studies running in parallel, estimate the number of units from the points you need to monitor at the same time, the turnaround between studies, and the spares you need. If your project involves 21 CFR Part 11 electronic record requirements, contact FRESVUE for information on the relevant LoggerAnalysis version and available validation documentation for your quality team’s review. FDA guidance on Part 11 scope and application